1、Designation:E104685(Reapproved 2015)Standard Specification forGlass Westergren Tube,Disposable1This standard is issued under the fixed designation E1046;the number immediately following the designation indicates the year oforiginal adoption or,in the case of revision,the year of last revision.A numb

2、er in parentheses indicates the year of last reapproval.Asuperscript epsilon()indicates an editorial change since the last revision or reapproval.1.Scope1.1 This specification covers a disposable tube used formeasuring the erythrocyte sedimentation rate,ESR(the sus-pension stability of red cells in

3、diluted,anti-coagulated humanblood).1.2 The values stated in SI units are to be regarded asstandard.No other units of measurement are included in thisstandard.2.Referenced Documents2.1 ASTM Standards:2E438 Specification for Glasses in Laboratory Apparatus3.Terminology3.1 Definitions of Terms Specifi

4、c to This Standard:3.1.1 disposable,nin accordance with this specificationand the expected product performance expressed in thisstandard,those tubes which are to be used one time only.Anyinstitution or individual who resuses a disposable tube mustbear full responsibility for its safety and effective

5、ness.3.1.2 Westergren3,nsurname of the individual responsiblefor the design of the reusable Westergren tube and the methodof use for both reusable and disposable tubes.4.Classification4.1 This specification covers a tube that is intended to beused one time only.It is not to be confused with a reusab

6、le tubethat is described in other published standards.The tubes shallbe of the following types:4.1.1 Type Istandard Westergren tube to be used in con-ventional racks only,and4.1.2 Type IIself-zeroing tubes to be used with racksdesigned with convenient disposable tube features.5.Materials5.1 GlassThe

7、 tubes made to this specification shall befabricated from borosilicate glass,Type I,Class B,or soda limeglass,Type II,in accordance with Specification E438.6.Physical Requirements6.1 Design,Type I and IIThe Westergren tube shall bemade of one-piece construction from tubing that is straight andof uni

8、form bore.The ends of the tube shall be cut at rightangles to the axis of the tube and fire polished.6.1.1 Type II onlyEach tube shall have an absorbant cottonplug inserted in the tube with the bottom edge of the plugaligned with the 0 mm calibration line.The alignment of theplug to the calibration

9、line should not vary by more than 1 mL.The plug must be capable of supporting a 200 mm verticalcolumn of water when the plug has been thoroughly soaked.NOTE1With possible printing scale variation and cotton plugalignment to the calibration line variation,the bottom of the cotton plug isto be located

10、 200 6 1 mm from the bottom of the tube.6.2 WorkmanshipThe tube shall be free as possible fromvisible defects which would detract from its appearance orimpair its serviceability when viewed by the human eye undernormal room lighting.Self-zeroing tubes are to be used withracks designed with convenien

11、t disposable tube features.6.3 DimensionsDimensions and tolerances shall be inaccordance with Fig.1.The uniformity of the bore shall be60.15 mm throughout the tube.The tube shall have aninscribed graduated scale extending over the lower 200 6 1.0mm of the tube.The markers or vendors name or mark sha

12、llbe inscribed on the tube in a selected location.Inscription ofthe word“Westergren”on the tube is optional.Type II tubesmay be made of varied lengths from 230 to 243 mm long withthe selected length within this range confined to a 61 mmtolerance.6.4 Graduation LinesThickness of graduation lines mayv

13、ary in uniformity but shall not exceed 0.4 mm in thickness.They shall lie at right angles to the axis of the tube with amaximum tolerance between two adjacent markings of 0.25mm.Maximum tolerance for the total 200 mm scale shall notexceed 1.0 mm.1This specification is under the jurisdiction of ASTM

14、Committee E41 onLaboratory Apparatus and is the direct responsibility of Subcommittee E41.01 onLaboratory Ware and Supplies.Current edition approved June 1,2015.Published July 2015.Originally approvedin 1985.Last previous edition approved in 2010 as E1046 85(2010).DOI:10.1520/E1046-85R15.2For refere

15、nced ASTM standards,visit the ASTM website,www.astm.org,orcontact ASTM Customer Service at serviceastm.org.For Annual Book of ASTMStandards volume information,refer to the standards Document Summary page onthe ASTM website.3Westergren,A.,“Studies of the Suspension Stability of the Blood in Pilmonary

16、Tuberculosis,”Acta Medica Scandinavia,54,19201921,pp.247282.Copyright ASTM International,100 Barr Harbor Drive,PO Box C700,West Conshohocken,PA 19428-2959.United States1 6.5 Graduation Line NumberingThe tube shall be gradu-ated in millimetres with a scale of 200 mm from the tip of thetube.The tube shall be numbered every 10 or 20 graduationlines starting with a numerical zero(0)and downward to amaximum value of 180 or 190 mm.The numerical markingsshall appear at the right side of the graduated s