收藏 分享(赏)

ASTM_E_1256_-_17.pdf

上传人:益****师 文档编号:168571 上传时间:2023-03-04 格式:PDF 页数:8 大小:273.14KB
下载 相关 举报
ASTM_E_1256_-_17.pdf_第1页
第1页 / 共8页
ASTM_E_1256_-_17.pdf_第2页
第2页 / 共8页
ASTM_E_1256_-_17.pdf_第3页
第3页 / 共8页
ASTM_E_1256_-_17.pdf_第4页
第4页 / 共8页
ASTM_E_1256_-_17.pdf_第5页
第5页 / 共8页
ASTM_E_1256_-_17.pdf_第6页
第6页 / 共8页
亲,该文档总共8页,到这儿已超出免费预览范围,如果喜欢就下载吧!
资源描述

1、Designation:E125617Standard Test Methods forRadiation Thermometers(Single Waveband Type)1This standard is issued under the fixed designation E1256;the number immediately following the designation indicates the year oforiginal adoption or,in the case of revision,the year of last revision.A number in

2、parentheses indicates the year of last reapproval.Asuperscript epsilon()indicates an editorial change since the last revision or reapproval.1.Scope1.1 The test methods described in these test methods can beutilized to evaluate the following six basic operational param-eters of a radiation thermomete

3、r(single waveband type):SectionCalibration Accuracy8Repeatability9Field-of-View10Response Time11Warm-Up Time12Long-Term Stability131.2 This standard does not purport to address all of thesafety concerns,if any,associated with its use.It is theresponsibility of the user of this standard to establish

4、appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.1.3 The term single waveband refers to radiation thermom-eters that operate in a single band of spectral radiation.Thisterm is used to differentiate single waveband radiation ther-mometer

5、s from those termed as ratio radiation thermometers,two channel radiation thermometers,two color radiationthermometers,multiwavelength radiation thermometers,mul-tichannel radiation thermometers,or multicolor radiation ther-mometers.The term single waveband does not precludewideband radiation thermo

6、meters such as those operating inthe 814 m band.1.4 This international standard was developed in accor-dance with internationally recognized principles on standard-ization established in the Decision on Principles for theDevelopment of International Standards,Guides and Recom-mendations issued by th

7、e World Trade Organization TechnicalBarriers to Trade(TBT)Committee.2.Referenced Documents2.1 ASTM Standards:E2758 Guide for Selection and Use of Wideband,LowTemperature Infrared Thermometers2.2 IEC DocumentsIEC/TS 62492-1 ed 1.0 TS Industrial Process ControlDevicesRadiation ThermometersPart 1:Techn

8、icalData for Radiation Thermometers3.Terminology3.1 Definitions:3.1.1 blackbody,nthe perfect or ideal source of thermalradiant power having a spectral distribution described by thePlanck equation.3.1.1.1 DiscussionThe term blackbody is often used todescribe a furnace or other source of radiant power

9、 whichapproximates the ideal.3.1.2 center wavelength,na wavelength,usually near themiddle of the band of radiant power over which a radiationthermometer responds,that is used to characterize its perfor-mance.3.1.2.1 DiscussionThe value of the center wavelength isusually specified by the manufacturer

10、 of the instrument.3.1.3 field-of-view,na usually circular,flat surface of ameasured object from which the radiation thermometer re-ceives radiation.2NOTE1Field-of-view traditionally has been referred to as target size.3.1.4 measuring distance,ndistance or distance rangebetween the radiation thermom

11、eter and the target(measuredobject)for which the radiation thermometer is designed.2NOTE2Measuring distance traditionally has been referred to as targetdistance.3.1.5 radiation thermometer,na radiometer calibrated toindicate the temperature of a blackbody.3.1.6 radiometer,na device for measuring rad

12、iant powerthat has an output proportional to the intensity of the inputpower.3.1.7 target distance,nsee measuring distance.3.1.8 target plane,nthe plane,perpendicular to the line ofsight of a radiation thermometer,that is in focus for thatinstrument.3.1.9 target size,nsee field-of-view.1These test m

13、ethods are under the jurisdiction of ASTM Committee E20 onTemperature Measurement and are the direct responsibility of Subcommittee E20.02on Radiation Thermometry.Current edition approved May 1,2017.Published June 2017.Originallyapproved in 1988.Last previous edition approved in 2015 as E1256 15.DOI

14、:10.1520/E1256-17.2IEC 62492-1.Copyright ASTM International,100 Barr Harbor Drive,PO Box C700,West Conshohocken,PA 19428-2959.United StatesThis international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles f

15、or theDevelopment of International Standards,Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade(TBT)Committee.1 3.2 Definitions of Terms Specific to This Standard:3.2.1 reference temperature source,na source of thermalradiant power of known temperature or e

16、missivity,or both,used in the testing of radiation thermometers.3.2.2 temperature resolution,nthe minimum simulated oractual change in target temperature that gives a usable changein output or indication,or both.4.Significance and Use4.1 The purpose of these test methods is to establishconsensus test methods by which both manufacturers and endusers may perform tests to establish the validity of the readingsof their radiation thermometers.The test results can also serveas standard performance cri

展开阅读全文
相关资源
猜你喜欢
相关搜索

当前位置:首页 > 专业资料 > 国外标准

copyright@ 2008-2023 wnwk.com网站版权所有

经营许可证编号:浙ICP备2024059924号-2