1、Designation:E127202(Reapproved 2012)Standard Specification for Laboratory GlassGraduated Cylinders1This standard is issued under the fixed designation E1272;the number immediately following the designation indicates the year oforiginal adoption or,in the case of revision,the year of last revision.A

2、number in parentheses indicates the year of last reapproval.Asuperscript epsilon()indicates an editorial change since the last revision or reapproval.1.Scope1.1 This specification covers requirements for glass gradu-ated cylinders for precision and general purpose grades suitablefor laboratory purpo

3、ses.1.1.1 Class AEach cylinder of precision grade shall bemarked with the letter“A”to signify compliance with appli-cable construction and accuracy requirements.Cylinders maybe marked with an identification number(serial number)at theoption of the manufacturer.1.1.2 Class BGeneral purpose cylinders

4、are of the samebasic design as Class A cylinders.However,volumetric toler-ances for Class B cylinders shall be within twice the specifiedrange allowed for Class A cylinders.These cylinders need notbe marked with their class designation.1.1.3 Product with a stated capacity not listed in thisstandard

5、may be specified in class A tolerance when productconforms to the tolerance range of the next smaller volumetricstandard product listed in Table 1.2.Referenced Documents2.1 ASTM Standards:2E438 Specification for Glasses in Laboratory ApparatusE542 Practice for Calibration of Laboratory VolumetricApp

6、aratusE675 Specification for Interchangeable Taper-Ground Stop-cocks And StoppersE920 Specification for Commercially Packaged LaboratoryApparatusE921 Specification for Export Packaged Laboratory Appa-ratusE1133 Practice for Performance Testing of Packaged Labo-ratory Apparatus for United States Gove

7、rnment Procure-mentsE1157 Specification for Sampling and Testing of ReusableLaboratory Glassware3.Classification3.1 Cylinders shall be in the following styles:3.1.1 Style IBeaded lip with pour spout,3.1.2 Style IIGround standard taper()neck,or3.1.3 Style IIIBeaded lip with pour spout and reinforcing

8、bead near top.4.Requirements4.1 Style I cylinders shall be calibrated either“to deliver”or“to contain.”Style II and Style III cylinders shall be calibrated“to contain.”The cylinders shall be calibrated at 20C and thevolumetric tolerances shall be in accordance with Table 1.(SeePractice E542 for cali

9、bration definitions and procedures.)4.1.1 Cylinders shall be of borosilicate glass conforming tothe requirements for Type I,Class A of Specification E438.4.2 ShapeCylinders shall be of one piece consisting ingeneral of a top,a graduated portion of uniform diameter anda base.The shape shall permit co

10、mplete emptying and thor-ough cleaning.The pour-out of Style I,in sizes of 1000 mLandbelow,shall be located in a vertical plane bisecting within 65one of the corners on the hexagonal base.For sizes 2000 mLand above,the pour-out shall bisect within 65 the center ofone of the flat section or one of th

11、e corners of the base.Thepour-out of Style III shall not protrude beyond the glassreinforcing bead with construction such that,upon falling,thebead shall strike a flat surface before the pour-out or lip.4.2.1 BaseThe base shall be hexagonal and the construc-tion shall be such that the empty cylinder

12、(without stopper forStyle II),in capacities above 10 mL,shall stand on a surfaceinclined 15 with the horizontal and,in capacities of 10 mL orless,shall stand on a surface inclined 10 with the horizontal.4.3 DimensionsThe dimensions shall be as shown inTable 1 The glass stoppers of Style II shall con

13、form toSpecification E675.4.4 Graduation MarkingsGraduation markings shall besharply defined lines of uniform width not more than 0.4 mm.Graduation markings shall be perpendicular to the vertical axisof the cylinder and parallel to each other.Adjacent graduationmarkings shall be spaced at least 1 mm

14、 from center to center.Markings shall be applied by one of the following methods:etched and filled with a permanent pigment;etched through a1This specification is under the jurisdiction of ASTM Committee E41 onLaboratory Apparatus and is the direct responsibility of Subcommittee E41.01 onApparatus.C

15、urrent edition approved Nov.1,2012.Published November 2012.Originallyapproved in 1989.Last previous edition approved in 2007 as E1272 02(2007).DOI:10.1520/E1272-02R12.2For referenced ASTM standards,visit the ASTM website,www.astm.org,orcontact ASTM Customer Service at serviceastm.org.For Annual Book

16、 of ASTMStandards volume information,refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International,100 Barr Harbor Drive,PO Box C700,West Conshohocken,PA 19428-2959.United States1 vertical colored band fired into the glass;or by application ofa stain which is fired into the glass without etching;or byapplication of enamel fired into the glass without etching;or byapplication of an enamel which is fired onto a vertical coloredetched band which has been fired into th