收藏 分享(赏)

ASTM_D_6978_-_05_2019.pdf

上传人:益****师 文档编号:170940 上传时间:2023-03-04 格式:PDF 页数:3 大小:86.28KB
下载 相关 举报
ASTM_D_6978_-_05_2019.pdf_第1页
第1页 / 共3页
ASTM_D_6978_-_05_2019.pdf_第2页
第2页 / 共3页
ASTM_D_6978_-_05_2019.pdf_第3页
第3页 / 共3页
亲,该文档总共3页,全部预览完了,如果喜欢就下载吧!
资源描述

1、Designation:D697805(Reapproved 2019)Standard Practice forAssessment of Resistance of Medical Gloves to Permeationby Chemotherapy Drugs1This standard is issued under the fixed designation D6978;the number immediately following the designation indicates the year oforiginal adoption or,in the case of r

2、evision,the year of last revision.A number in parentheses indicates the year of last reapproval.Asuperscript epsilon()indicates an editorial change since the last revision or reapproval.1.Scope1.1 This practice covers a protocol for the assessment ofresistance of medical glove materials to permeatio

3、n by poten-tially hazardous cancer chemotherapy drugs under conditionsof continuous contact.An assessment is made based on thepermeation(breakthrough)of nine chemotherapy drugsthrough the glove material over a certain period of time.1.2 It is emphasized that the conditions used in this assess-ment a

4、re intended to approximate the worst-case condition forclinical uses.The data should be restricted to use on a relativebasis when comparing glove materials.1.3 This standard does not purport to address all of thesafety concerns,if any,associated with its use.It is theresponsibility of the user of th

5、is standard to establish appro-priate safety,health,and environmental practices and deter-mine the applicability of regulatory limitations prior to use.1.4 This international standard was developed in accor-dance with internationally recognized principles on standard-ization established in the Decis

6、ion on Principles for theDevelopment of International Standards,Guides and Recom-mendations issued by the World Trade Organization TechnicalBarriers to Trade(TBT)Committee.2.Referenced Documents2.1 ASTM Standards:2D471 Test Method for Rubber PropertyEffect of LiquidsD3577 Specification for Rubber Su

7、rgical GlovesD3578 Specification for Rubber Examination GlovesD3767 Practice for RubberMeasurement of DimensionsD5250 Specification for Poly(vinyl chloride)Gloves forMedical ApplicationD6319 Specification for Nitrile Examination Gloves forMedical ApplicationD6977 Specification for Polychloroprene Ex

8、aminationGloves for Medical ApplicationF739 Test Method for Permeation of Liquids and GasesThrough Protective Clothing Materials Under Conditionsof Continuous Contact3.Terminology3.1 Definitions:3.1.1 See Test Method F739,Section 3 on Terminology fora full list of terms and definitions.3.1.2 breakth

9、rough detection time,nthe time in minutesmeasured from the start of the test to the sampling time thatimmediately precedes the sampling time at which the perme-ation rate reaches 0.01 g/cm2/min.See Test Method F739,section 12.4.2 for calculation of permeation rate with replen-ishment.3.1.3 validated

10、 statistical rationale,n requires that anobjective cause or reason be stated as to why the initial test wasincorrect.4.Significance and Use4.1 The objective of this practice is to provide a uniformprocedure for assessing the resistance of medical glove mate-rials to permeation by chemotherapy drugs,

11、and to establish aconsistent reporting of the test data.5.Test Protocol5.1 Summary of Test Method:5.1.1 The gloves in question shall be tested in accordancewith the method set out in Test Method F739,Procedure A.5.1.2 The resistance of a medical glove material to perme-ation by chemotherapy drugs sh

12、all be determined by measuringthe breakthrough detection time of the drugs through the glovematerial.The test method involves using the test material serveas a membrane(partition)between the two halves of a test cell.One half of the cell shall contain the donor solution of the testdrug and the other

13、 half shall contain the appropriate collectionmedium,that is,distilled water or other liquid,which does notinfluence the permeation of the drug being tested.A test drugpermeates through the sample material when the test drug isable to pass through the test material.As the test drug1This practice is

14、under the jurisdiction ofASTM Committee D11 on Rubber andRubber-like Materials and is the direct responsibility of Subcommittee D11.40 onConsumer Rubber Products.Current edition approved May 1,2019.Published June 2019.Originallyapproved in 2005.Last previous edition approved in 2013 as D6978 05(2013

15、).DOI:10.1520/D6978-05R19.2For referenced ASTM standards,visit the ASTM website,www.astm.org,orcontact ASTM Customer Service at serviceastm.org.For Annual Book of ASTMStandards volume information,refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International,100 Barr Ha

16、rbor Drive,PO Box C700,West Conshohocken,PA 19428-2959.United StatesThis international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for theDevelopment of International Standards,Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade(TBT)Committee.1 continuously passes through the sample material,its concen-tration increases over time.The collection medium shall b

展开阅读全文
相关资源
猜你喜欢
相关搜索

当前位置:首页 > 专业资料 > 国外标准

copyright@ 2008-2023 wnwk.com网站版权所有

经营许可证编号:浙ICP备2024059924号-2