1、Designation:D515106(Reapproved 2015)Standard Test Method forDetection of Holes in Medical Gloves1This standard is issued under the fixed designation D5151;the number immediately following the designation indicates the year oforiginal adoption or,in the case of revision,the year of last revision.A nu
2、mber in parentheses indicates the year of last reapproval.Asuperscript epsilon()indicates an editorial change since the last revision or reapproval.1.Scope1.1 This test method covers the detection of holes inmedical gloves.1.2 This test method is limited to the detection of holes thatallow water lea
3、kage under the conditions of the test.1.3 The smallest hole size that will allow water leakage in amedical glove has not been determined and is beyond the scopeof this test method.1.4 The safe and proper use of medical gloves is beyond thescope of this test method.1.5 The values stated in SI units a
4、re to be regarded asstandard.No other units of measurement are included in thisstandard.1.6 This standard does not purport to address all of thesafety concerns,if any,associated with its use.It is theresponsibility of the user of this standard to establish appro-priate safety and health practices an
5、d determine the applica-bility of regulatory limitations prior to use.2.Terminology2.1 Definitions of Terms Specific to This Standard:2.1.1 medical gloves,nmedical gloves,as used in this testmethod,refer to both surgical and examination gloves.3.Significance and Use3.1 This test method is designed t
6、o detect holes that leakwater and thereby compromise the usefulness of the glove.3.2 This test method is suitable and designed as a referencemethod to evaluate samples of medical gloves.4.Apparatus4.1 MandrelA clear plastic tube,nominally 6 6 0.5 cmoutside diameter,5 cm inside diameter or 50 mm sche
7、dule 80poly(vinyl chloride)(PVC)pipe,38 6 1 cm long.The mandrelshall have a clamp or rubber band for securing the glove to themandrel.4.2 Graduated Cylinder or BeakerAs a minimum,thegraduated cylinder or beaker shall be marked in increments of50 cm3to a capacity of 1000 cm3.The markings shall beaccu
8、rate to within 610 cm3at volumes greater than 500 cm3.4.3 Stopwatch.4.4 Various Clamps,Hangers,and Stands.5.Reagents5.1 Tap WaterThe water used in this test method shall beclean and free from any visible contaminants.6.Procedure6.1 Throughout 6.2 and 6.3,ensure that the exterior of theglove remains
9、dry.6.2 Mount the mandrel in a vertical position using appro-priate stands,clamps,and hangers.Affix the glove to themandrel by stretching the cuff of the glove around the mandrel.A maximum 40 mm of glove cuff should fit over the bottomend of the mandrel.Use the securing device,as necessary,tohold th
10、e glove in place.The remainder of the glove shouldhang freely from the mandrel when filled with water.6.3 Pour a minimum of 1000 cm3of water having a roomtemperature of 15 to 30C into the top of the mandrel.Thewater shall pass freely into the glove.6.4 Visually inspect the glove for immediate water
11、leakage.Let the glove hang for 2 min and again inspect for waterleakage.7.Interpretation of Results7.1 Any glove that shows a droplet,stream,or other type ofwater leakage shall be considered to have failed the test.7.2 Water leakage above the junction of the glove cuff andmandrel is not a test failu
12、re.If this leakage occurs,remove anddry the outside of the glove and retest.8.Report8.1 The report shall include the number of samples tested,the number of failures,and a general description of the locationof each hole.1This test method is under the jurisdiction of ASTM Committee D11 on Rubberand is
13、 the direct responsibility of Subcommittee D11.40 on Consumer RubberProducts.Current edition approved Nov.1,2015.Published December 2015.Originallyapproved in 1990.Last previous edition approved in 2011 as D5151 06(2011).DOI:10.1520/D5151-06R15.Copyright ASTM International,100 Barr Harbor Drive,PO B
14、ox C700,West Conshohocken,PA 19428-2959.United States1 9.Precision and Bias9.1 No statement is made about either the precision or biasof this test method.This is a“pass-fail”test procedure notsubject to the usual analysis for quantitative variable precision.10.Keywords10.1 detection;gloves;holes;med
15、icalASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentionedin this standard.Users of this standard are expressly advised that determination of the validity of any such patent rights,and the riskof infringement of such rights,ar
16、e entirely their own responsibility.This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years andif not revised,either reapproved or withdrawn.Your comments are invited either for revision of this standard or for additional standardsand should be addressed to ASTM International Headquarters.Your comments will receive careful consideration at a meeting of theresponsible technical committee,which you may attend.If you feel that yo