1、Designation:D51114(Reapproved 2021)1Standard Test Methods forCalcium and Magnesium In Water1This standard is issued under the fixed designation D511;the number immediately following the designation indicates the year oforiginal adoption or,in the case of revision,the year of last revision.A number i
2、n parentheses indicates the year of last reapproval.Asuperscript epsilon()indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the U.S.Department of Defense.1NOTEThe WTO caveat was added editorially in December 2021.1.Scope1.1 The
3、se test methods cover the determination of calciumand magnesium in water by complexometric titration andatomic absorption spectrometric procedures.Two test methodsare included,as follows:SectionsTest Method AComplexometric Titration7 15Test Method BAtomic Absorption Spectrometric16 251.2 The values
4、stated in SI units are to be regarded asstandard.The values given in parentheses are mathematicalconversion to inch-pound units that are provided for informa-tion only and are not considered standard.1.3 This standard does not purport to address all of thesafety concerns,if any,associated with its u
5、se.It is theresponsibility of the user of this standard to establish appro-priate safety,health,and environmental practices and deter-mine the applicability of regulatory limitations prior to use.Specific hazard statements are given in 12.2.6 and 20.6.1.4 This international standard was developed in
6、 accor-dance with internationally recognized principles on standard-ization established in the Decision on Principles for theDevelopment of International Standards,Guides and Recom-mendations issued by the World Trade Organization TechnicalBarriers to Trade(TBT)Committee.2.Referenced Documents2.1 AS
7、TM Standards:2D1129 Terminology Relating to WaterD1193 Specification for Reagent WaterD2777 Practice for Determination of Precision and Bias ofApplicable Test Methods of Committee D19 on WaterD3370 Practices for Sampling Water from Flowing ProcessStreamsD4691 Practice for Measuring Elements in Water
8、 by FlameAtomic Absorption SpectrophotometryD4841 Practice for Estimation of Holding Time for WaterSamples Containing Organic and Inorganic ConstituentsD5810 Guide for Spiking into Aqueous SamplesD5847 Practice for Writing Quality Control Specificationsfor Standard Test Methods for Water Analysis3.T
9、erminology3.1 DefinitionsFor definitions of terms used in these testmethods,refer to Terminology D1129.4.Significance and Use4.1 Calcium and magnesium salts in water are the primarycomponents of water hardness which can cause pipe or tubescaling.5.Purity of Reagents5.1 Reagent grade chemicals shall
10、be used in all tests.Unless otherwise indicated,it is intended that all reagents shallconform to the specifications of the Committee on AnalyticalReagents of the American Chemical Society.3Other gradesmay be used,provided it is first ascertained that the reagent isof sufficiently high purity to perm
11、it its use without lesseningthe accuracy of the determination.5.2 Purity of WaterUnless otherwise indicated,referencesto water shall be understood to mean reagent water conformingto Specification D1193,Type 1,II,or III water.Type I ispreferred and more commonly used.Type II water wasspecified at the
12、 time of round-robin testing of these testmethods.NOTE1The user must ensure the type of reagent water is sufficiently1These test methods are under the jurisdiction of ASTM Committee D19 onWater and are the direct responsibility of Subcommittee D19.05 on InorganicConstituents in Water.Current edition
13、 approved Nov.1,2021.Published December 2021.Originallyapproved in 1937.Last previous edition approved in 2014 as D511 14.DOI:10.1520/D0511-14R21E01.2For referenced ASTM standards,visit the ASTM website,www.astm.org,orcontact ASTM Customer Service at serviceastm.org.For Annual Book of ASTMStandards
14、volume information,refer to the standards Document Summary page onthe ASTM website.3Reagent Chemicals,American Chemical Society Specifications,AmericanChemical Society,Washington,DC.For Suggestions on the testing of reagents notlisted by the American Chemical Society,see Annual Standards for Laborat
15、oryChemicals,BDH Ltd.,Poole,Dorset,U.K.,and the United States Pharmacopeiaand National Formulary,U.S.Pharmacopeial Convention,Inc.(USPC),Rockville,MD.Copyright ASTM International,100 Barr Harbor Drive,PO Box C700,West Conshohocken,PA 19428-2959.United StatesThis international standard was developed
16、in accordance with internationally recognized principles on standardization established in the Decision on Principles for theDevelopment of International Standards,Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade(TBT)Committee.1?free of interferences.The water should be analyzed using this test method.6.Sampling6.1 Collect the sample in accordance with Practices D3370.6.2 If total recoverable calcium and magnesium concentra-tions are being determined,