1、Designation:D 3797 05Standard Test Method forAnalysis of o-Xylene by Gas Chromatography1This standard is issued under the fixed designation D 3797;the number immediately following the designation indicates the year oforiginal adoption or,in the case of revision,the year of last revision.A number in

2、parentheses indicates the year of last reapproval.Asuperscript epsilon(e)indicates an editorial change since the last revision or reapproval.1.Scope*1.1 This test method covers the analysis of normally occur-ring impurities in,and the purity of,o-xylene by gas chroma-tography.Impurities determined i

3、nclude nonaromatic hydro-carbons,benzene,toluene,p-and m-xylenes,cumene,styrene,and ethylbenzene.1.2 This test method is applicable for impurities at concen-trations from 0.001 to 2.000%and for o-xylene purities of98%or higher.1.3 The following applies to all specified limits in thisstandard:for pur

4、poses of determining conformance with thisstandard,an observed value or a calculated value shall berounded off“to the nearest unit”in the last right-hand digitused in expressing the specification limit,in accordance withthe rounding-off method of Practice E 29.1.4 The values stated in SI units are t

5、o be regarded asstandard.No other units of measurement are included in thisstandard.1.5 This standard does not purport to address all of thesafety concerns,if any,associated with its use.It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and de

6、termine the applica-bility of regulatory limitations prior to use.For a specifichazard statement,see Section 8.2.Referenced Documents2.1 ASTM Standards:2D 3437 Practice for Sampling and Handling Liquid CyclicProductsD 4307 Practice for Preparation of Liquid Blends for Use asAnalytical StandardsD 479

7、0 Terminology of Aromatic Hydrocarbons and Re-lated ChemicalsD 6809 Guide for Quality Control and Quality AssuranceProcedures for Aromatic Hydrocarbons and Related Ma-terialsE 29 Practice for Using Significant Digits in Test Data toDetermine Conformance with SpecificationsE 355 Practice for Gas Chro

8、matography Terms and Rela-tionshipsE 1510 Practice for Installing Fused Silica Open TubularCapillary Columns in Gas Chromatographs2.2Other Documents:OSHA Regulations,29 CFR,paragraphs 1910.1000 and1910.120033.Terminology3.1 Definitions:3.1.1 For definition of terms used in this test method seeTermin

9、ology D 4790.4.Summary of Test Method4.1 A known amount of internal standard is added to thesample.Agas chromatograph equipped with a flame ionizationdetector and a polar-fused silica capillary column is used forthe analysis.The impurities are measured relative to theinternal standard.To calculate o

10、-Xylene purity subtract theimpurities found from 100.00%.5.Significance and Use5.1 This test method is suitable for setting specifications ono-xylene and for use as an internal quality control tool whereo-xylene is used in a manufacturing process.It may be used indevelopment or research work involvi

11、ng o-xylene.5.2 Purity is commonly reported by subtracting the deter-mined expected impurities from 100%.Absolute purity cannotbe determined if unknown impurities are present.6.Apparatus6.1 Gas ChromatographyAny gas chromatograph havinga flame ionization detector and a splitter injector suitable for

12、use with a fused-silica capillary column may be used,providedthe system has sufficient sensitivity to obtain a minimum peak1This test method is under the jurisdiction of ASTM Committee D16 onAromatic Hydrocarbons and Related Chemicals and is the direct responsibility ofSubcommittee D16.01 on Benzene

13、,Toluene,Xylenes,Cyclohexane and TheirDerivatives.Current edition approved July 1,2005.Published August 2005.Originallyapproved in 1979.Last previous edition approved in 2004 as D 3797 04.2For referenced ASTM standards,visit the ASTM website,www.astm.org,orcontact ASTM Customer Service at serviceast

14、m.org.For Annual Book of ASTMStandards volume information,refer to the standards Document Summary page onthe ASTM website.3Available from U.S.Government Printing Office Superintendent of Documents,732 N.Capitol St.,NW,Mail Stop:SDE,Washington,DC 20401.1*A Summary of Changes section appears at the en

15、d of this standard.Copyright ASTM International,100 Barr Harbor Drive,PO Box C700,West Conshohocken,PA 19428-2959,United States.height response of 0.1 mV for 0.010%internal standard whenoperated at the stated conditions.Background noise at theseconditions is not to exceed 3 V.6.2 Chromatographic Col

16、umn,fused silica capillary,60-mlong,0.32-mm inside diameter,internally coated to a 0.5-mthickness with a bonded(cross-linked)polyethylene glycol.Other columns may be used after it has been established thatsuch column is capable of separating all major impurities andthe internal standard from the o-xylene under operating con-ditions appropriate for the column.6.3 Recorder,optional.6.4 Electronic integration with tangent capabilities(re-quired).6.5 Microsyringes,10-L,and 50-L.6.6 Volumetric Flask,