收藏 分享(赏)

ASTM_D_4127_-_18a.pdf

上传人:益****师 文档编号:181270 上传时间:2023-03-04 格式:PDF 页数:7 大小:122.17KB
下载 相关 举报
ASTM_D_4127_-_18a.pdf_第1页
第1页 / 共7页
ASTM_D_4127_-_18a.pdf_第2页
第2页 / 共7页
ASTM_D_4127_-_18a.pdf_第3页
第3页 / 共7页
ASTM_D_4127_-_18a.pdf_第4页
第4页 / 共7页
ASTM_D_4127_-_18a.pdf_第5页
第5页 / 共7页
ASTM_D_4127_-_18a.pdf_第6页
第6页 / 共7页
亲,该文档总共7页,到这儿已超出免费预览范围,如果喜欢就下载吧!
资源描述

1、Designation:D412718aStandard Terminology Used withIon-Selective Electrodes1This standard is issued under the fixed designation D4127;the number immediately following the designation indicates the year oforiginal adoption or,in the case of revision,the year of last revision.A number in parentheses in

2、dicates the year of last reapproval.Asuperscript epsilon()indicates an editorial change since the last revision or reapproval.1.Scope*1.1 This terminology covers those terms recommended bythe International Union of Pure and Applied Chemistry(IUPAC),2and is intended to provide guidance in the use ofi

3、on-selective electrodes for analytical measurement of speciesin water,wastewater,and brines.1.2 This international standard was developed in accor-dance with internationally recognized principles on standard-ization established in the Decision on Principles for theDevelopment of International Standa

4、rds,Guides and Recom-mendations issued by the World Trade Organization TechnicalBarriers to Trade(TBT)Committee.2.Referenced Documents2.1 ASTM Standards:3D1129 Terminology Relating to Water3.Terminology3.1 Definitions:3.1.1 For definitions of terms used in this standard,refer toTerminology D1129.3.2

5、 Definitions of Terms Relevant to Ion-Selective ElectrodeTechnology:acid error,nin very acid solutions,the activity of water isreduced(less than unity)causing a non-Nernstian responsein glass electrodes.DISCUSSIONA positive error in the pH reading results.activity,nthe thermodynamically effective co

6、ncentration ofa free ion in solution.DISCUSSIONIn dilute solutions,ionic activity,and concentration arepractically identical,but in solutions of high ionic strength,or in thepresence of complexing agents,activity may differ significantly fromconcentration.Ionic activity,not concentration,determines

7、both therate and the extent of chemical reactions.activity coefficient,na factor,that relates activity,A,to theconcentration,C of a species in solution:A 5 CDISCUSSIONThe activity coefficient is dependent on the ionicstrength of the solution.Ions of similar size and charge have similaractivity coeff

8、icients.alkaline error,nin alkaline solutions,where hydrogen ionactivity becomes very small,some glass electrodes respondto other cations,such as sodium.DISCUSSIONAnegative error in the pH reading results.By changingthe composition of the glass,the affinity of the glass for sodium ion canbe reduced.

9、Such electrodes are known as lithium glass,high-pH,orfull-range electrodes.analate,nthe sample being analyzed;used in the terms“analate addition”and“analate subtraction.”DISCUSSIONThis term differs from the term“analyte,”whichdescribes the chemical species of interest in an analytical test.asymmetry

10、 potential,nthe potential across a glass pHelectrode membrane when the inside and outside of themembrane are in contact with solutions of identical pH.DISCUSSIONThis term has also been used to define the observedpotential differences between identical electrode pairs placed in iden-tical solutions.c

11、alibration curve,na plot of the potential(emf or E)of agiven ion-selective electrode cell assembly(ion-selectiveelectrode combined with an identified reference electrode)versus the logarithm of the ionic activity(or concentration)of a given species.DISCUSSIONFor uniformity,it is recommended that the

12、 potential beplotted on the ordinate(vertical axis)with the more positive potentialsat the top of the graph and that paA(log activity of the speciesmeasured,A)or pcA(log concentration of species measured,A)beplotted on the abscissa(horizontal axis)with increasing activity to theright.See Fig.1 and F

13、ig.2.Region I of Fig.1 represents the segmentof the curve where the potential no longer changes in response tochanges of the measured species activity and the electrode no longerdemonstrates Nernstian response.IUPACactivity standard,na standardizing solution whose value isreported in terms of ionic

14、activity.DISCUSSION If the electrode is calibrated using activity standards,the activity of the free,unbound ion in the sample is determined.1This terminology is under the jurisdiction of ASTM Committee D19 on Waterand is the direct responsibility of Subcommittee D19.05 on Inorganic Constituentsin W

15、ater.Current edition approved May 1,2018.Published June 2018.Originallyapproved in 1982.Last previous edition approved in 2018 as D4127 18.DOI:10.1520/D4127-18a.2Recommendations for Nomenclature of Ion-Selective Electrodes,IUPAC Com-mission on Analytical Nomenclature,Pergamon Press,Oxford,1976.3For

16、referenced ASTM standards,visit the ASTM website,www.astm.org,orcontact ASTM Customer Service at serviceastm.org.For Annual Book of ASTMStandards volume information,refer to the standards Document Summary page onthe ASTM website.*A Summary of Changes section appears at the end of this standardCopyright ASTM International,100 Barr Harbor Drive,PO Box C700,West Conshohocken,PA 19428-2959.United StatesThis international standard was developed in accordance with internationally recognized principles

展开阅读全文
相关资源
猜你喜欢
相关搜索

当前位置:首页 > 专业资料 > 国外标准

copyright@ 2008-2023 wnwk.com网站版权所有

经营许可证编号:浙ICP备2024059924号-2