1、Designation:F 1205 88(Reapproved 1999)Standard Specification forAnesthesia Breathing Tubes1This standard is issued under the fixed designation F 1205;the number immediately following the designation indicates the year oforiginal adoption or,in the case of revision,the year of last revision.A number

2、in parentheses indicates the year of last reapproval.Asuperscript epsilon(e)indicates an editorial change since the last revision or reapproval.1.Scope1.1 This specification covers anesthesia breathing tubes,integrally attached end fittings,and integrally attached compo-nents for use with anesthesia

3、 breathing systems.Excluded aresimilar tubes for specific other purposes,such as those usedwith some breathing machines,pediatric anesthesia circuits,and gas scavenging tubes.1.2 The following subjects are covered:terminology,di-mensions,labeling and marking,selected performance require-ments,and te

4、st methods to substantiate compliance with therequirements.Alternative test methods may be employedprovided that they can be shown to be equivalent or better.NOTE1This specification supersedes ANSI Standard Z 79.6-1975.1.3 This standard does not purport to address all of thesafety problems,if any,as

5、sociated with its use.It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2.Referenced Documents2.1 ASTM Standards:F 1054 Specification For Conical Fittings Of 15 mm and 22m

6、m Sizes22.2ANSI Standard:ANSI/NFPA 99 Healthcare Facilities33.Terminology3.1 Descriptions of TermsFor the purpose of this specifi-cation,the following descriptions of terms shall apply:3.1.1 ambient temperature and pressureroom temperatureand pressure,dry gas(rtpd);indicates conditions of 20C and101

7、.3 kPa(760 mm Hg)and dry gas.3.1.2 assembled enda fitting integrally attached by themanufacturer to the end of the tube.3.1.3 breathing tubea large-bore(greater than or equal to15 mm)nonrigid tube,usually corrugated,used to conveygases or vapors,or both,to and from the patient.3.1.4 compliance(press

8、ure-volume relationship)a changeof volume per change in pressure.Units of compliance areexpressed as millilitres per kilopascal(millilitres per centime-tre H2O)per metre length of tube.3.1.5 conductivea property of a breathing tube and anyintegrally attached components with electrical conductivityac

9、ceptable in accordance with the applicable test(see 10.1.1).3.1.6 disposablean item intended to be used once anddiscarded.3.1.7 flow resistancethe pressure difference from inlet tooutlet of a device caused by flow.Units of resistance arekilopascal per litre per second(centimetre of H2O per litre per

10、second)per meter length of tube.3.1.8 labelinginformation and literature accompanyingthe device,such as brochures,package inserts,and manuals.3.1.9 machine endthat end of the breathing tube which istoward the machine.3.1.10 markinginformation directly on the device.3.1.11 nonconductivea property of

11、a breathing tube andany integrally attached components with insufficient electricalconductivity to meet the requirements of applicable test(see10.1.1).3.1.12 patient endthat end of the breathing tube which istoward the patient.3.1.13 reusablean item intended for repeated use on oneor more patients.4

12、.Material4.1 Requirements:4.1.1 Tubes should be made of materials suitable for theintended use and should function satisfactorily in the presenceof anesthetic agents and gases commonly used and not elutetoxic substances.4.1.2 Reusable tubing should be able to withstand ordinarymethods of cleaning,di

13、sinfection,and sterilization(see12.1.1.1).It is desirable that such products should withstandaccepted methods of steam sterilization.1This specification is under the jurisdiction of ASTM Committee F29 onAnesthetic and Respiratory Equipment and is the direct responsibility of Subcom-mittee F29.12 on

14、Airways.Current edition approved Dec.30,1988.Published February 1989.2Annual Book of ASTM Standards,Vol 13.01.3Available from American National Standards Institute,25 W.43rd St.,4thFloor,New York,NY 10036.1Copyright ASTM International,100 Barr Harbor Drive,PO Box C700,West Conshohocken,PA 19428-2959

15、,United States.5.Dimensions and Tolerances5.1 Requirements:5.1.1 Measure SystemSI units shall be used.5.1.2 End Connectors:5.1.2.1 Single Breathing Tube(s)Both ends of the tubeshall mate with 22-mm male fittings which comply withSpecification F 1054.The ends may be assembled,plain,cylindrical,or tap

16、ered.5.1.2.2 Assembled Sets(integrally attached Y-piece andtubing,not intended to be disassembled):(a)Patient EndThe patient end of the Y-piece shall be a22-mm/15-mm male/female coaxial fitting complying withSpecification F 1054.(b)Machine EndThe machine end of both breathing tubesshall mate with 22-mm male fittings which comply withSpecification F 1054.The ends may be assembled,plain,cylindrical,or tapered.5.1.3 LengthThe length of the breathing tube shall bedesignated as the nominal overall le