收藏 分享(赏)

ASTM_F_564_-_10_2015.pdf

上传人:益****师 文档编号:190059 上传时间:2023-03-04 格式:PDF 页数:13 大小:364.10KB
下载 相关 举报
ASTM_F_564_-_10_2015.pdf_第1页
第1页 / 共13页
ASTM_F_564_-_10_2015.pdf_第2页
第2页 / 共13页
ASTM_F_564_-_10_2015.pdf_第3页
第3页 / 共13页
ASTM_F_564_-_10_2015.pdf_第4页
第4页 / 共13页
ASTM_F_564_-_10_2015.pdf_第5页
第5页 / 共13页
ASTM_F_564_-_10_2015.pdf_第6页
第6页 / 共13页
亲,该文档总共13页,到这儿已超出免费预览范围,如果喜欢就下载吧!
资源描述

1、Designation:F56410(Reapproved 2015)Standard Specification and Test Methods forMetallic Bone Staples1This standard is issued under the fixed designation F564;the number immediately following the designation indicates the year of originaladoption or,in the case of revision,the year of last revision.An

2、umber in parentheses indicates the year of last reapproval.Asuperscriptepsilon()indicates an editorial change since the last revision or reapproval.1.Scope1.1 This specification covers characterization of the designand mechanical function of metallic staples used in the internalfixation of the muscu

3、lar skeletal system.It is not the intentionof this specification to describe or specify specific designs formetallic bone staples.1.2 This specification includes the following four test meth-ods for measuring mechanical properties of metallic bonestaples:1.2.1 Test Method for Constant Amplitude Bend

4、ing FatigueTests of Metallic Bone StaplesAnnex A1.1.2.2 Test Method for Pull-Out Fixation Strength of Metal-lic Bone StaplesAnnex A2.1.2.3 Test Method for Soft Tissue Fixation Strength ofMetallic Bone StaplesAnnex A3.1.2.4 Test Method for Elastic Static Bending of MetallicBone StaplesAnnex A4.1.3 Th

5、e values stated in SI units are to be regarded asstandard.No other units of measurement are included in thisstandard.1.4 This standard does not purport to address all of thesafety concerns,if any,associated with its use.It is theresponsibility of the user of this standard to establish appro-priate s

6、afety and health practices and determine the applica-bility of regulatory limitations prior to use.2.Referenced Documents2.1 ASTM Standards:2E4 Practices for Force Verification of Testing MachinesE122 Practice for Calculating Sample Size to Estimate,WithSpecified Precision,the Average for a Characte

7、ristic of aLot or ProcessE467 Practice for Verification of Constant Amplitude Dy-namic Forces in an Axial Fatigue Testing SystemF75 Specification for Cobalt-28 Chromium-6 MolybdenumAlloy Castings and Casting Alloy for Surgical Implants(UNS R30075)F86 Practice for Surface Preparation and Marking of M

8、etal-lic Surgical ImplantsF382 Specification and Test Method for Metallic Bone PlatesF565 Practice for Care and Handling of Orthopedic Implantsand InstrumentsF601 Practice for Fluorescent Penetrant Inspection of Me-tallic Surgical ImplantsF629 Practice for Radiography of Cast Metallic SurgicalImplan

9、ts3.Finish3.1 Staples conforming to this specification shall be finishedand identified in accordance with Practice F86,as appropriate.4.Inspection Practices4.1 Staples made in accordance with Specification F75should be inspected in accordance with Practice F601 orX-rayed in accordance with Practice

10、F629.5.Care and Handling5.1 Staples should be cared for and handled in accordancewith Practice F565,as appropriate.6.Keywords6.1 bending test;bone fixation;fatigue test;fixation devices;metallic bone staples;orthopaedic medical devices;pullouttest;soft tissue fixation;surgical implants1This specific

11、ation is under the jurisdiction of ASTM Committee F04 onMedical and Surgical Materials and Devices and is the direct responsibility ofSubcommittee F04.21 on Osteosynthesis.Current edition approved March 1,2015.Published April 2015.Originallyapproved in 1985.Last previous edition approved in 2010 as

12、F564 10.DOI:10.1520/F0564-10R15.2For referenced ASTM standards,visit the ASTM website,www.astm.org,orcontact ASTM Customer Service at serviceastm.org.For Annual Book of ASTMStandards volume information,refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International,100 B

13、arr Harbor Drive,PO Box C700,West Conshohocken,PA 19428-2959.United States1 ANNEXES(Mandatory Information)A1.TEST METHOD FOR CONSTANT AMPLITUDE BENDING FATIGUE TESTS OF METALLIC BONE STAPLESA1.1.ScopeA1.1.1 This test method covers procedures for the perfor-mance of constant amplitude fatigue testing

14、 of metallic staplesused in internal fixation of the musculoskeletal system.Thistest method may be used when testing in air at ambienttemperature or in an aqueous or physiological solution.A1.1.2 The values stated in SI units are to be regarded asstandard.No other units of measurement are included i

15、n thisstandard.A1.1.3 This standard does not purport to address all of thesafety concerns,if any,associated with its use.It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.

16、A1.2.Summary of Test MethodA1.2.1 Metallic bone staples are tested under bending loadsuntil the specimen fails or a predetermined number of cycleshas been applied to it.Bending tests may be performed in oneof two modes:either pure,in-plane bending;or tension(orcompression)combined with in-plane bending.Tests usingeither of these methods may be conducted at ambient condi-tions or in aqueous or physiological solutions(at either roomtemperature or 37C).A1.3.Significance and UseA1.3.1 This test meth

展开阅读全文
相关资源
猜你喜欢
相关搜索

当前位置:首页 > 专业资料 > 国外标准

copyright@ 2008-2023 wnwk.com网站版权所有

经营许可证编号:浙ICP备2024059924号-2