1、Designation:F 965 99Standard Specification forRigid Laryngoscopes for Tracheal IntubationHook-OnFittings for Laryngoscope Handles and Blades with Lamps1This standard is issued under the fixed designation F 965;the number immediately following the designation indicates the year oforiginal adoption or
2、,in the case of revision,the year of last revision.A number in parentheses indicates the year of last reapproval.Asuperscript epsilon(e)indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the Department of Defense.1.Scope1.1 This
3、 specification covers critical dimensions and con-siderations for the junction of any blade and any handle of ahook-on type laryngoscope to allow engagement,lamp illumi-nation,and disengagement through multiple cycles.An elec-trical lamp is fixed to the blade and supplied with energythrough the hand
4、le.1.2 This standard does not cover:(1)the blade form orhandle design beyond the interchangeability aspects of theconnection between the blade and the handle;(2)the measure-ment and specificiation of the lamp illumination intensity;and(3)disposable,flexible,or fiber-illuminated laryngoscopesdesigned
5、 for surgery.1.3 The state-of-the-art of laryngoscope performance doesnot permit sufficient reproducibility to allow inclusion ofmandatory force values in this standard;therefore,only rec-ommended values have been provided.1.4 The values stated in SI units are to be regarded as thestandard.2.Referen
6、ced Documents2.1ANSI Standard:ANSI Y14.5 Dimensions and Tolerances23.Terminology3.1 Definitions of Terms Specific to This Standard:3.1.1 bladea separate rigid device to provide direct visionof the larynx,incorporating a hook-on base and a socketconnection for a laryngoscope lamp and an electrical co
7、nnec-tion to the hook-on end.3.1.2 contactsconductive components of the hook-on fit-tings,both handle and blade,which come together to make anelectrical circuit when the handle and blade are in the locked,operating position.3.1.3 engagementthe blade and handle are engaged whenthe hinge pin is positi
8、oned against the closed end of the hingeslot,and remains coupled in all positions.3.1.4 folded positionthe position of the engaged bladewhen the tip of the blade is as close as possible to the handle.3.1.5 handlea separate,rigid device which is held in thehand during use,containing the hook-on conne
9、ction thatretains the blade and supplies electrical energy for the lamp.3.1.6 hingethe mechanical feature of the handle and bladethat allows them to be engaged and disengaged.The hinge alsoallows the blade to be placed into the operating,folded,and allintermediate positions while engaged.3.1.7 hinge
10、 pinthe component of the handle that insertsinto the blade hinge slot.3.1.8 hinge slotthe surfaces of the blade hook-on fittingthat engage the hinge pin.3.1.9 locking mechanismthe feature that retains the bladehook-on fitting in the operating position.3.1.10 locking slotsthe surfaces of the handle h
11、ook-onfitting that accept the locking surfaces of the blade hook-onfitting.3.1.11 locking surfacesthe components of the blade lock-ing mechanism that fit into the locking slots to resist the bladedisplacement from the normal operating position.3.1.12 operating positionthe position of the engagedblad
12、e and handle when the lamp is lit and the instrument isready for use.3.1.13 retainerthe surface,component,or feature of theblade hinge slot that retains the blade hook-on fitting to thehandle hinge pin.The retainer maintains the engagement of theblade in the operating,folded,and all intermediate pos
13、itions.3.1.14 seating surfacesthe surface or surfaces of the bladehook-on fitting that contact the handle hook-on fitting tosupport the blade in the locked operating position.4.Significance and Use4.1 The primary objective of this specification is to ensureinterchangeability concerning mechanical fi
14、t and electricalcontact between the hook-on type handle and laryngoscopeblade by specifying only the critical dimensions of the handle1This specification is under the jurisdiction of ASTM Committee F29 onAnesthetic and Respiratory Equipment and is the direct responsibility of Subcom-mittee F29.13 on
15、 Laryngoscopes and Bronchoscopes.Current edition approved July 10,1999.Published September 1999.Originallypublished as F 965 85.Last previous edition F 965 85(1993)e1.2Available from American National Standards Institute(ANSI),25 W.43rd St.,4th Floor,New York,NY 10036.1Copyright ASTM International,1
16、00 Barr Harbor Drive,PO Box C700,West Conshohocken,PA 19428-2959,United States.hook-on fitting that affect interchangeability,and by designat-ing the minimum performance requirements.4.2 The secondary objective is to provide test proceduresand values for performance.5.Performance Requirements5.1 Hook-on Blade and Handle FittingsDetachablehook-on blade and handle combinations that engage shall lockand the lamp will light when in the operating position.NOTE1For rationale statements for the perform