1、Mars 2006 March 2006 CEI 61223-3-5(Premire dition 2004)IEC 61223-3-5(First edition 2004)Essais dvaluation et de routine dans les services dimagerie mdicale Partie 3-5:Essais dacceptation-Performance dimagerie des quipements de tomodensitomtrie rayonnement X Evaluation and routine testing in medical
2、imaging departments Part 3-5:Acceptance tests Imaging performance of computed tomography X-ray equipment CORRIGENDUM 1 Page 56 F.4 Evaluation des donnes avec le DISPOSITIF DESSAI alternatif mire barres parallles Remplacer lquation existante:eaumatriaufondmireio2CTCTNMMMFTM=par la nouvelle quation su
3、ivante:eaumatriau2fond2mireio2CTCTNMMMFTM=Page 57 F.4 Data evaluation with the alternate TEST DEVICE“bar pattern”Replace the existing equation:watermaterialbackgroundpatternio2CTCTNMMMMTF=with the following new equation:watermaterial2background2patternio2CTCTNMMMMTF=Page 62 Bibliographie Au lieu de:
4、8 JUDY,P.F.The line spread function and modulation transfer function of a computed tomographic scanner.J Comput Assist Tomogr 3,1976,p.189-195.lire:8 JUDY,P.F.The line spread function and modulation transfer function of a computed tomographic scanner.Medical Physics,July,1976,Vol.3,Issue 4,p.233-236
5、 Page 63 Bibliography Instead of:8 JUDY,P.F.The line spread function and modulation transfer function of a computed tomographic scanner.J Comput Assist Tomogr 3,1976,p.189-195.read:8 JUDY,P.F.The line spread function and modulation transfer function of a computed tomographic scanner.Medical Physics,July,1976,Vol.3,Issue 4,p.233-236