1、February 2012 IEC 60601-2-20(Second edition 2009)Medical electrical equipment Part 2-20:Particular requirements for the basic safety and essential performance of infant transport incubators C O R R I G E N D U M 1 201.4.3.101 ESSENTIAL PERFORMANCE of INFANT TRANSPORT INCUBATORS Replace the existing
2、Table 201.101 by the following:Table 201.101 Additional ESSENTIAL PERFORMANCE requirements Requirement Subclause ESSENTIAL PERFORMANCE requirement 1 201.12.1.104 or generation of a visual and audible alarm in compliance with 201.9.6.2.1.102 ESSENTIAL PERFORMANCE requirement 2 201.12.1.106 or generat
3、ion of a visual and audible alarm in compliance with 201.9.6.2.1.102 201.9.6.2.1.101 Sound level within the COMPARTMENT In the first paragraph,replace the reference to 201.12.3.107 by 201.9.6.2.1.103.Fvrier 2012 CEI 60601-2-20(Deuxime dition 2009)Appareils lectromdicaux Partie 2-20:Exigences particu
4、lires pour la scurit de base et les performances essentielles des incubateurs de transport pour nouveau-ns C O R R I G E N D U M 1 201.4.3.101 Performances essentielles des incubateurs de transport pour nouveau-ns Remplacer le Tableau 201.101 existant par le nouveau tableau suivant:Tableau 201.101 E
5、xigences supplmentaires de PERFORMANCES ESSENTIELLES Exigence Paragraphe Exigence 1 de PERFORMANCES ESSENTIELLES 201.12.1.104 ou gnration dune alarme visuelle et sonore conforme 201.9.6.2.1.102 Exigence 2 de PERFORMANCES ESSENTIELLES 201.12.1.106 ou gnration dune alarme visuelle et sonore conforme 201.9.6.2.1.102 201.9.6.2.1.101 Niveau de pression acoustique lintrieur du COMPARTIMENT Dans le premier alina,remplacer la rfrence 201.12.3.107 par 201.9.6.2.1.103.