1、February 2013 Fvrier 2013 IEC 60601-2-20 (Second edition 2009)Medical electrical equipment Part 2-20:Particular requirements for the basic safety and essential performance of infant transport incubators CEI 60601-2-20(Deuxime dition 2009)Appareils lectromdicaux Partie 2-20:Exigences particulires pou

2、r la scurit de base et les performances essentielles des incubateurs de transport pour nouveau-ns C O R R I G E N D U M 2 201.9.6.2.1.102*Audible alarms sound level Replace the existing first sentence by the following:Auditory ALARM SIGNALS shall have a sound level of at least 65 dBA at a distance o

3、f 3 m perpendicular to the front of the control unit in a reflecting room.201.9.6.2.1.102*Niveau de pression acoustique de lalarme sonore Remplacer la premire phrase existante par la suivante:Dans un local rflchissant,les SIGNAUX DALARME sonore doivent produire un niveau de pression acoustique dau moins 65 dBA une distance de 3 m perpendiculaire lavant de llment de commande.