1、 November 2012-1-IEC 60601-1(Edition 3.1 2012)Consolidated version Medical electrical equipment Part 1:General requirements for basic safety and essential performance CORRIGENDUM 1 7.9.3 Technical description 7.9.3.1 General Delete the first bullet in the list that immediately follows the first para
2、graph of this subclause,which reads:the information required in 7.2;Table D.2 Safety signs For rows 5,6 and 7,instead of:5 ISO 7010-P017 Pushing prohibited 6 ISO 7010-P018 Sitting prohibited 7 ISO 7010-P019 Stepping prohibited November 2012-2-read:5 ISO 7010-P017 No pushing 6 ISO 7010-P018 No sitting 7 ISO 7010-P019 No stepping on surface