收藏 分享(赏)

ASTM_E_1079_-_10.pdf

上传人:益****师 文档编号:168438 上传时间:2023-03-04 格式:PDF 页数:2 大小:66.22KB
下载 相关 举报
ASTM_E_1079_-_10.pdf_第1页
第1页 / 共2页
ASTM_E_1079_-_10.pdf_第2页
第2页 / 共2页
亲,该文档总共2页,全部预览完了,如果喜欢就下载吧!
资源描述

1、Designation:E107910Standard Practice forCalibration of Transmission Densitometers1This standard is issued under the fixed designation E1079;the number immediately following the designation indicates the year oforiginal adoption or,in the case of revision,the year of last revision.A number in parenth

2、eses indicates the year of last reapproval.Asuperscript epsilon()indicates an editorial change since the last revision or reapproval.1.Scope1.1 This practice2covers the calibration of transmissiondensitometers used to perform radiographic film density mea-surements(see Note 1).1.2 This standard does

3、 not purport to address all of thesafety concerns,if any,associated with its use.It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.NOTE1For further information on the desi

4、gn and use ofdensitometers,the following literature is suggested as additional back-ground information:ISO 51:2009,ISO 52:2009,and ISO 14807:2001.2.Referenced Documents2.1 ASTM Standards:3E1316 Terminology for Nondestructive Examinations2.2 ISO Standards:4ISO 51:2009 Photography and graphic technolo

5、gy-Den-sity measurements-Part 1:Geometry and functionalnotationISO 52:2009 Photography and graphic technology-Den-sity measurements-Part 2:Geometric conditions fortransmittance densityISO 14807:2001 Photography-Transmission and reflectiondensitometers-Method for determining performance3.Terminology3

6、.1 DefinitionsFor definitions of terms used in thispractice,see Terminology E1316.4.Significance and Use4.1 This practice provides a means for calibrating transmis-sion densitometers used for the measurement of radiographicfilm density.A transmission densitometer calibrated in accor-dance with this

7、practice provides the assurance that accuratedensity values of radiographs are obtained.5.Apparatus5.1 Apparatus should consist of the following:5.1.1 A calibrated step tablet shall be used.The step tabletmay be a NIST X-ray Step Tablet(X-Ray Film Step TabletTransmission Density Standard 38100C)5or

8、alternately a steptablet from another supplier,which is traceable to the NISTstep tablet.The step tablet shall have at least five step densities,which cover the density range that is used for productionradiographs.Acalibration certificate shall be provided with thestep tablet indicating the tablet I

9、D and recorded values for eachstep density.For suppliers of step tablets other than NIST,thecertificate shall indicate conformance of traceability to NISTinstrumentation used in the calibration process,applicableANSI standards used,verification of measurement on a NISTstep tablet,the ID number of th

10、e step tablet,and calibrationdate of the step tablet.Precautions should be taken in thestorage,handling,and use of the step tablet.In the event itbecomes scratched,blemished,or exhibits other signs ofdeleterious wear,it should be replaced immediately.The NIST(or alternate,if used)step tablet shall b

11、e replaced four yearsfrom the date of first use.65.1.2 Transmission Densitometers,with either direct-scalereadout or digital readout displays specifically manufacturedfor the purpose of measuring the range of film densitiesdescribed in 5.1.1 may be used.5.1.3 Manufacturers Operating Instructions for

12、 Appropri-ate Transmission Densitometer.6.Calibration6.1 Full-scale linearity calibration should be performed atleast every 90 days during use as follows:1This practice is under the jurisdiction of ASTM Committee E07 on Nonde-structive Testing and is the direct responsibility of Subcommittee E07.01

13、onRadiology(X and Gamma)Method.Current edition approved June 1,2010.Published July 2010.Originally approvedin 1985.Last previous edition approved in 2005 as E1079-05.DOI:10.1520/E1079-10.2For ASME Boiler and Pressure Vessel Code applications see related PracticeSE-1079 in Section II of that Code.3Fo

14、r referenced ASTM standards,visit the ASTM website,www.astm.org,orcontact ASTM Customer Service at serviceastm.org.For Annual Book of ASTMStandards volume information,refer to the standards Document Summary page onthe ASTM website.4Available from American National Standards Institute(ANSI),25 W.43rd

15、 St.,4th Floor,New York,NY 10036.5Available from National Institute of Standards and Technology(NIST),100Bureau Dr.,Stop 1070,Gaithersburg,MD 20899-1070,http:/www.nist.gov.6Expiration interval of the NIST or alternate step tablet may be different than therequirements of this practice.Unless otherwis

16、e specified,requirements of thispractice shall apply.Copyright ASTM International,100 Barr Harbor Drive,PO Box C700,West Conshohocken,PA 19428-2959.United States1 6.1.1 Use the manufacturers recommended warm-up timeto stabilize circuitry before starting the calibration procedure orthe periodic verification checks described in Section 7.Adjustthe“0”reading of the densitometer after the warm-up period.6.1.2 Select and measure three steps on the calibrated steptablet densities below,above,and near

展开阅读全文
相关资源
猜你喜欢
相关搜索

当前位置:首页 > 专业资料 > 国外标准

copyright@ 2008-2023 wnwk.com网站版权所有

经营许可证编号:浙ICP备2024059924号-2