1、Designation:D226893(Reapproved 2017)Standard Test Method forAnalysis of High-Purity n-Heptane and Isooctane byCapillary Gas Chromatography1This standard is issued under the fixed designation D2268;the number immediately following the designation indicates the year oforiginal adoption or,in the case

2、of revision,the year of last revision.A number in parentheses indicates the year of last reapproval.Asuperscript epsilon()indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the U.S.Department of Defense.1.Scope1.1 This test meth

3、od covers and provides for the analysis ofhigh-purity(greater than 99.5%by volume)n-heptane andisooctane(2,2,4-trimethylpentane),which are used as primaryreference standards in determining the octane number of a fuel.Individual compounds present in concentrations of less than0.01%can be detected.Col

4、umns specified by this test methodmay not allow separation of all impurities in reference fuels.1.2 The values stated in SI units are to be regarded as thestandard.1.2.1 ExceptionThe values given in parentheses are forinformation only.1.3 This standard does not purport to address all of thesafety co

5、ncerns,if any,associated with its use.It is theresponsibility of the user of this standard to establish appro-priate safety,health,and environmental practices and deter-mine the applicability of regulatory limitations prior to use.1.4 This international standard was developed in accor-dance with int

6、ernationally recognized principles on standard-ization established in the Decision on Principles for theDevelopment of International Standards,Guides and Recom-mendations issued by the World Trade Organization TechnicalBarriers to Trade(TBT)Committee.2.Summary of Test Method2.1 The sample is injecte

7、d into a capillary gas chromato-graphic column consisting of at least 61 m(200 ft)of stainlesssteel tubing(0.25 mm(0.010 in.)inside diameter),the innerwalls of which are coated with a thin film of stationary liquid.An inert gas transports the sample through the column,inwhich it is partitioned into

8、its individual components.As eachcomponent is eluted from the column,it is detected with ahydrogen flame ionization detector and recorded on a conven-tional strip-chart recording potentiometer.The detector re-sponse from each impurity is then compared with that of aknown quantity of an internal stan

9、dard.After determining thetotal impurity concentration,the n-heptane,or isooctane purityis obtained by difference.3.Significance and Use3.1 This test method is used for specification analysis ofhigh-purity n-heptane and isooctane,which are used as ASTMKnock Test Reference Fuels.Hydrocarbon impuritie

10、s orcontaminants,which can adversely affect the octane number ofthese fuels,are precisely determined by this method.4.Apparatus4.1 ChromatographGaschromatographshouldbeequipped with a split-stream inlet device for introducingminute quantities of sample without fractionation,a capillarycolumn,and a h

11、ydrogen flame ionization detector.An elec-trometer to amplify the low output signal of the hydrogen flameionization detector,and a strip-chart recorder for recording thedetector signal are needed.The time constant of neither theelectrometer nor the recorder should exceed 1 s.Aball and diskintegrator

12、 or electronic integrator for peak area measurementsshould be used.The detection system must have sufficientsensitivity to produce a recorder deflection for cyclohexane ofat least 8 divisions on a standard 0100 scale chart using0.10%by volume of cyclohexane in n-heptane as defined in7.1.4.2 Microsyr

13、ingeA microsyringe is needed for injectingthe sample into the split-stream inlet device.4.3 Volumetric Pipet,0.1 mL capacity.4.4 Analytical Balance,200 g capacity.5.Reagents and Materials5.1 Carrier GasArgon,Nitrogen,or Helium;99.99%orgreater purity.(WarningCompressed gases under high pres-sure.)5.2

14、 Fuel GasHydrogen;99.99%or greater purity.(WarningCompressed gas under high pressure.Extremelyflammable gas.)1This test method is under the jurisdiction of ASTM Committee D02 onPetroleum Products,Liquid Fuels,and Lubricants and is the direct responsibility ofSubcommittee D02.04.0L on Gas Chromatogra

15、phy Methods.Current edition approved Oct.1,2017.Published November 2017.Originallyapproved in 1964.Last previous edition approved in 2013 as D2268 93(2013).DOI:10.1520/D2268-93R17.Copyright ASTM International,100 Barr Harbor Drive,PO Box C700,West Conshohocken,PA 19428-2959.United StatesThis interna

16、tional standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for theDevelopment of International Standards,Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade(TBT)Committee.1 5.3 OxidantGasAir;99.99%orgreaterpurity.(WarningCompressed gases under high pressure.)5.4 CyclohexaneAt least 99 mol%pure,to be used asinternal standard.(WarningFlammable liquid and harmful ifinge